Control valve: To carry out all functions of the cycle at one point, i.e. Jacket to Chamber, Chamber fast/slow exhaust, Vacuum drying.
Door: Single door fitted with radial arms & automatic pressure locking device so that the door cannot be opened until the steam is fully exhausted from the chamber.
Quick vacuum drying apparatus: Allows filtered and sterile air to break the chamber vacuum and helps in quick drying.
Accidental vacuum breaker / jacket air valve: A safety device for the jacket against accidental vacuum and for automatic air removal to obtain saturated steam.
Safety valve: A safety device against excess pressure in jacket / steam generator. Dial type thermometer indicates sterilization temperature.
Non-return valve: Swing type valves, S.S. 304 quality.
Chamber condensate line: Fitted with a thermostatic steam trap and check valve for thorough evacuation of air and condensate from the chamber to achieve optimum temperature, and an extra by-pass line in case of emergency.
Drain line: S.S. 304 quality, 2" N.B. drain.
Steam generator (boiler): Attached underneath the jacket.
M.O.C: S.S. 304 quality, 14 G thick, with heater plate of S.S. 304 quality 12 mm thick, and boiler cover with perforation.
Heater: Water immersion type, industrial, 09 kW total.
Water level switch: Magnetic switch to protect heater from low water level.
Glass tube: To indicate water level in boiler.
S.S. ball valves: For water inlet & outlet.
Inside electrical panel: Contactor: Siemens make.
Pressure switch: Danfoss make.
Indicating lamp: Neon lamp for main and heater.
Key features
Fully stainless steel, panel boiler cover & hinge
Digital temperature indicator cum controller cum timer & alarm
Low water level cut off system
In-built steam generator
Automation
An embedded programmable controller is a digital computer used for automation of electromechanical processes, such as control of machinery.
Cycles: Sterilization of varied equipment such as gloves, linen, gauze, cotton, instruments, glassware etc., using standard 121°C & 134°C cycles which are pre-programmed in the microprocessor.
No. of pulses, vacuum strength for pulse, sterilization time, operating temperature & pressure, drying time & post vacuum strength: all these parameters are selectable & programmable as per requirement.
Fully automatic operation.
Self-diagnostic system is incorporated in microprocessor programming and provides fault / error message with indication & alarm on system failure. (1) Control: During sterilization, the embedded controller controls the temperature, moisture, valve control, and all critical time and pressure-based parameters.
(2) Temperature is maintained uniformly & accurately (± 0.5°C). Continuous display of temperature, pressure, and humidity on-line allows the user to be aware of cycle status at all times.
(3) Alarm: In case of any error, an error code & alarm is generated and the cycle stops automatically.
(4) We hope that you will find the above quotation in line with your requirements. Please feel free to contact us for any further clarifications. It will be our pleasure to assist you the best.
(5) We hope that our quotation is most competitive and in line with your requirement. Thanking you and waiting for your valued order at the earliest.
Technical specifications
Horizontal Cylindrical Steam Sterilizer, Single Door, Electrically operated, Standard Model
Operating pressure: 1.2 to 2.3 kg/cm²
Operating temperature: 121°C to 134°C
Partial Vacuum: 10" Hg
Full Vacuum: 26" Hg, through vacuum pump only (Optional)
Hydro test: Jacket: 2.6 kgf/cm² g; Chamber: 1.9 kgf/cm² g